MACView-Particles measurement case
The MACView®Particles Measurement Case did arise to meet the demand for mobile measurements. Both the Solvent-version and the Particles-versie can be integrated in a case. A MACView®-Particles Measurement Case is applicable for various things, like giving a demonstration at a prospect or another relation, or to do measurements on various locations in big production halls. Combined with a laptop, you can directly create reports of the measurements made.
Thorough and mobile
As you can see on the pictures below, this version of the MACView®-Particles Measurement Case gets delivered in an solid case, which gives sufficient protection to be used in an industrial environment. A MACView®-Particles Measurement Case is standard delivered with 1,5 meter cable for the sensor and another 1,5 meter for a connection to a PC or laptop. Of course, this MACView® product is standart delivered with the software.
For true mobile measurements we offer the option to build an battery in the case. By doing so, you are not dependent on a nearby plug-socket and/or a extension cable. Connected to this battery, the MACView®-Particles Measurement Case is able to do measurements during 20 hours.
Dimensions case | 460 x 440 x 160 mm |
Approval | CE |
Protection classification read-out unit | IP65 |
Standard case contains | - 1 MACView® (particles- or solvent-version) - 1 MACView® sensor (particles- or solvent-version) - 1,5 meter kabel for the sensor - 1,5 meter kabel for PC-connection - 1,5 meter kabel for connection to mains socket The MACView®-Case is ‘ready-to-run’! |
Material sensor housing | highgrade stainless steel |
Probe length (solvent-version) | 192 mm |
Sensor diameter (particles-version) | 130 mm |
Warming-up time | 120 sec. for MACView® solvent 240 sec. for MACView® particles |
Short-term stability | - solvent-version: 1% - particles-version: ± 3% |
Short-term stability > 1 year | - solvent-version: ± 2% - particles-version: ± 5% |
Signaling / Alarming | - solvent-version: ppm and hysteresis adjustable per relais - particles-version: mg/m3 and hysteresis adjustable per relais |
Displaying | using back-lit LCD display with 2 rows x 16 characters |
Units MAC-value on display | - solvent-version: ppm - particles-version: mg/m3 |
Acknowledgement alarm | by use of reset pushbutton on the MACView® |
Communication to PC | via an RS232-cable |
Speed of the microcontroller | up to 20 MHz |
Fastening | Read-out unit and sensor can be fastened to the wall together or apart easily. |
Software | The MACregister® software is useable for Windows 95, 98, ME, NT, 2000 and Windows XP. |
- Operation principle particles sensor
- Leaflet MACView-Koffer
- Verwarming en ventilatie 2003V6
- Application note MACView-Particles V2.0 Measurement Case
- Achtergrondinformatie en testrapporten bij de MACView®-Particles Measurement Case
- Application note MACView-Particles (English)
- Intelligent sensors control the air quality on the workplace (English)
- Leaflet MACView-Measurement Case (English)
- Operation principle particles sensor (English)
- Gebruiksaanwijzing_Particles-NL (Dutch)